- Main PLC controls & power disconnect installed in NEMA 12 enclosure(s) under framework. Enclosure meets NEC, CSA, UL, CE. Main PLC & power controls NOT mounted on skirting or guards as this does not meet the above certifications requiring a rated enclosure.
- Mitsubishi FX3u PLC for all process controls.
- Beijer 10” color touch screen operator interface terminal (OIT-HMI); stores multiple recipes, requires ID/PW sign-in.
- On-screen programmable Time and Temperature controls w/alarm limits is standard.
- Individualized operator, supervisor/engineer or calibration unique sign in ID/PW for assigned access levels by administrator.
- USB port (powered) for output of Audit data or input of barcodes with hand held scanner (not provided) on each stroke.
- Odometer and resettable cucle counter
- Data Logging memory of all sign-on access, changes and UDI process data per each cycle. Memory accessible for manual offloading only. No edits permitted.
- Capability to match recipe woth barcoded tooling and barcoded device or paperwork during setup of production.
- Redundant automatic over temperature safety cutoff control circuit, Factory Mutual approved controller, 300F cutoff standard.
- Electronic process monitoring with temperature & air pressure sensor fail-safe alarms for parameter recipe settings, PW protected. All alarms are logged.
- Four (4) dual lead special limits grade thermocouples per ASTM E230-12 table 1 for accurate control feedback & individual temperature displays on OIT while providing signals to calibration ports for simultaneous reading, PW protected.
- Alarm screen detail history list on screen and in memory.
- Low profile Banner multi-color signal light for seal parameter & process alarms.
- Pressure regulation w/precision manual adjustment & liquid filled precision gauge in addition to digital readout.
- Pressure validation port for easy calibration gage connection to pressure circuit. On-screen digital calibration of pressure circuit, PW protected.
- Individual thermocouple calibration connectors. On-screen calibration of each (4) special grade dual channel thermocouples, PW protected.
- Timer calibration connector (1).
- Eye level controls including 10” color touch OIT/HMI screen, reset(s) and others.
- English or metric units of measure readouts (specify at ordering)
- Air Usage 11-13 CFM @ 80 PSI at 4 cycles per minute. Supply @ 100psi with 3/4”+ dia line min.
- Single exhaust port with 20 micron filter; or for connection to discharge line from room
Owner’s manual (1 copy) with operation instructions, maintenance information, recommended spares list, schematics, Atlas Vac calibration document and Atlas Vac PLC programs on CD ROM. Sealer is ready for operation upon connection to correct air and electric power; air filter included. Any additional FAT, or other documents and procedures, may require additional charge based on scope and hours needed to complete.
Tooling nests not included, quoted separately.