- Mitsubishi PLC FX3u
- Beijer 10” iX T10F-2 color touch screen operator interface terminal (OIT-HMI); stores multiple recipes, requires ID/PW sign-in for Operators.
- Barcode reader port (powered) for input, ex. UDI, Operator PW.
- On-screen programmable Time and Temperature controls w/alarm limits is standard
- Individualized operator, supervisor/engineer or calibration unique sign in ID/PW for assigned access levels by administrator.
- Odometer and resettable cycle counter.
- USB port (powered) for output of data or input of barcodes with hand held scanner (not provided) . Keyboard capable also.
- Data Logging memory of all sign-on access, changes and UDI as well as process data per each cycle. Memory accessible for manual offloading (standard) or optional network upload. No edits permitted.
- Capability to match screen recipe with barcoded tooling and barcoded device/paperwork during setup of production.
- Redundant automatic heater over temperature safety cutoff control circuit, Factory Mutual approved controller, 300F cutoff standard
- Electronic process monitoring with temperature & air pressure sensor fail-safe alarms for parameter recipe settings, PW protected. All alarms are logged.
- Four (4) dual lead special limits grade thermocouples per ASTM E230-12 table 1 for accurate control feedback & individual temperature displays on OIT while providing signals to calibration ports for simultaneous reading.
- Individual thermocouple calibration connectors (4). On screen calibration of each (4) thermocouples. PW protected.
- Alarm screen detail history list on screen and in memory.
- Low profile Banner multi-color signal light for seal parameter & process alarms.
- Pressure regulation w/precision manual adjustment & liquid filled precision gauge in addition to digital readout on screen
- Pressure validation port for easy calibration gage connection to pressure circuit. On-screen digital calibration of pressure circuit, PW protected.
- Timer calibration connector (1),
- Eye level controls including 10” color touch OIT/HMI screen, reset(s) and others.
- English units of measure readouts.
- Air Usage 11-13 CFM @ 80 PSI at 4 cycles per minute. Supply @ 100psi with 3/4”+ diameter line minimum,
- Single exhaust port with 20 micron filter; or for connection to discharge line from room
Owner’s manual (1 copy) with operation instructions, maintenance information, recommended spares list, schematics, Atlas Vac calibration document and Atlas Vac PLC programs on USB memory stick. Sealer is ready for operation upon connection to correct air and electric power; air filter included. Any additional FAT, or other documents and procedures, may require additional charge based on scope and hours needed to complete.
Tooling nests not included, quoted separately.